NUVE Benchtop Clinical Sterilizers

  • Both B & S type of models are available.
  • Chamber Capacity:  NC 23B-23S : 23 liters, NC 32S : 32 liters.
  • Conforms to the requirements of EN 13060+A2 standard.
  • Powerful and reliable N-Smart™ microprocessor control system.
  • Massive data storage with internal memory which stores the records of last 500 cycles in detail and 29.500 cycles as summary.
  • Memory for the last 100 failures.

Overview

With enormous experience in the manufacturing of steam sterilizers. 

With enormous experience in the manufacturing of steam sterilizers,this current product range has been carefully designed to cover all applications in general medical practices, dentistry, facilities for personal hygiene, beauty care, veterinary practices and stand-by use in hospitals and clinics. Instruments conform to the requirements of the EN 13060 standard for B and S Class sterilization with additional technical advantages.

Elimination of air bubbles is important and steam should reach all surfaces, so B and S units have a fractionated vacuum operation with a vacuum pump designed especially for use with steam. S models, because of the types of loads that will be sterilized, do not need such a feature and use fractionated gravity with alternating inflow and outflow of steam.

Traceability is comprehensively covered with storage and visualization of all run data using N-Smart™ control system. A large range of safety features and self-diagnostics provide full user protection.

product range has been carefully designed to cover all applications in general medical practices, dentistry, facilities for personal hygiene, beauty care, veterinary practices and stand-by use in hospitals and clinics. Instruments conform to the requirements of the EN 13060 standard for B and S Class sterilization with additional technical advantages.

Elimination of air bubbles is important and steam should reach all surfaces, so B and S units have a fractionated vacuum operation with a vacuum pump designed especially for use with steam. S models, because of the types of loads that will be sterilized, do not need such a feature and use fractionated gravity with alternating inflow and outflow of steam.

Traceability is comprehensively covered with storage and visualization of all run data using N-Smart™ control system. A large range of safety features and self-diagnostics provide full user protection.

  • Type of Sterilization Loads:
    B Models: All wrapped, un-wrapped, solid, hollow load products, Type A* and porous products
    S models: Non-wrapped solid products, porous products, small porous items, hollow load products Type B*, single wrapped products, multiple-layer wrapped products
  • Sterilization programs:
    B Models: 5 pre-set programs(Universal, quick unwrapped material, sensitive material, prion, flash)
    S Models: 5 pre-set programs (Quick, universal, prion, sensitive, wrapped)
  • 2 free programs and 1 drying program.
  • Test Programs:
    B models: Bowie & Dick / Helix Test and Vacuum Test
    S models: Bowie & Dick and Vacuum Test.
  • Operating parameters can be tracked graphically against time.
  • USB port to record the operated cycles in the memory to a memory stick.
  • Ethernet port for remote access through internet by means of optional NuveCloser™ software.
  • Optional panel type printer to keep the records of the sterilization cycle.
  • Password protected control system, service and calibration menu.
  • Lid safety system.
  • Color of the display changes to red in case of any alarm condition along with an audible alarm.
  • Reminder for gasket and filter replacement.
  • Remote failure diagnostics by means of NuveCloser™ software.
  • Sending e-mails up to five e-mail addresses with the details of failure.
  • Optional AlerText™ SMS alarm system for ultimate security.
  • Safety valve against overpressure.
  • HEPA filter at air intake line.
  • Certified chamber according to PED 2014/68 EU.
  • Fractionated pre-vacuum system
  • Process evaluation system.
  • Comprehensive self-diagnostic system.

* As defined in EN 13060 +A2 Standard.

Specifications

 

Model Nos.

NC 23B

NC 23S

NC 32S

Chamber Volume (liters)

23

23

32

Sterilization Temperature

121°C and 134°C

Number of Preset Programs

5

Number of Free Programs

2

Test Programs

Vacuum leakage , Bowie & Dick/Helix

Vacuum leakage , Bowie & Dick

Minimum Vacuum Level

100 mbar

300 mbar

Control System

N-Smart™ Control System

Display

4,3” Colorful LCD Display

Chamber Material

Stainless Steel

Number Of Shelves (Standard/max) pcs

3/6

3/8

Capacity For Sterilization Cassettes, pcs

3

Power Supply

220 V - 50/60 Hz

Power Consumption

2.500 W

3.000 W

Chamber Dimensions (Diameter x Depth) mm

Ø 260x440

Ø 320x400

External Dimensions (WxDxH) mm

570 x 730 x 525

625 x 710 x 575

Packing Dimensions (WxDxH) mm

770 x 630 x 690

740 x 680 x 750

Net/Packed Weight kg

55 / 69

65 / 82

Video

Variations

Image Product Name Catalog No. Product Description Pack Size Quantity Action

Accessories

Cat. No.

Description    

Qty/Pack

FACTORY FITTED OPTIONS    

NC XXX Y    

Panel type thermal printer

 

each

A 08 191       

AlerText™ GSM alarm module

each

A 08 212      

NuveCloser™ Software CD with 3 m. RS 232 cable

each

R 01 123       

Aluminum shelf for NC 23B/S

each

R 01 125       

Aluminum shelf for NC 32S

each

A 08 092      

Sealed sterilization pack carrier NC 23B/S

each

A 08 112       

Sealed sterilization pack carrier NC 32/S

each